Redox measuring amplifier
|
|
Measuring range: |
-1500 mV to +1500 mV |
Resolution: |
1 mV |
Ambient temperature influence: |
50 ppm/K (12 mV) |
Reference temperature: |
23° C |
Input resistance: |
> 1012 ohm |
Redox measurement: |
via combination electrode without reference system |
Calibration: |
not required |
pH measuring amplifier
|
|
Measuring range: |
-1500 mV to +1500 mV |
Resolution: |
0.01 pH or 1 mV |
Ambient temperature influence: |
50 ppm/K (12 mV) |
Reference temperature: |
23° C |
Input resistance: |
1012 ohm |
Calibration pH: |
optional single-point or two-point calibration |
Electrode check: |
automatically after calibration |
Temperature compensation of pH value: |
automatic via PT1000 or manually adjustable |
Temperature
|
|
Temperature input for measuring sensor: |
PT1000 |
Measuring range: |
-30°C to +140°C (+/-1%) |
Inputs/outputs, control response
|
|
Type of control: |
P, PI, or PID control optionally |
Control: |
optionally 1-position |
Binary outputs: |
2 potential-free output relays for disinfection, pH reduce/pH raise (pulse length, pulse frequency, On/Off), alarm |
Maximum contact load (230V/AC): |
370 W (inductive), 500 W (Ohm resistive load) |
Max. switching current: |
30V/10A, 115V/0,3A, 230V/0,12A |
Digital inputs: |
Empty signal chlorine, pH reduce, dosing stop |
Environmental conditions
|
|
Working temperature: |
5 to +50 C° |
Storage temperature: |
20 to 65 C° |
Rel. humidity: |
max. 90 % at 40 °C (non-condensing) |